NanoDrop ND-1000 3.7. NanoDrop ND-1000. Choose the most popular programs from Audio & Video software. 3.7.1 ( See all) NanoDrop Technologies. Review Comments Questions & Answers Update program info. Protocol for use of NanoDrop® ND-1000 For Nucleic acids 1. Turn on the software by clicking on the icon. (Note: there is no ON/OFF switch on the instrument) Choose your “User” folder for storage of your data. The first panel allows for the selection of the required analysis and the “Nucleic Acid” button is on the top left of the. Open the software of the NanoDrop by double clicking at the icon 'ND-1000 V.3.2.1' on the desktop. To measure the protein concentration at 280 nm press the 'Protein A280' button. The programme will ask to place 2 μl of deionized water (dH 2 O) on the lower sample pedestal and then press OK. In 'Sample type' box choose '1 Abs = 1 mg/ml'. NanoDrop 1000 Software is required to operate the NanoDrop 1000 instrument. This software logs the data received from the NanoDrop 1000 in an archive file on the computer. The sample data can be opened by a spreadsheet program such as Microsoft Excel. NanoDrop 2000 1.3. NanoDrop 2000 software allows you to interact with the NanoDrop 2000 and NanoDrop 2000c devices. Besides interaction, NanoDrop 2000 controls and manages all the features and the modules accessible through the same interface.

Overview

NanoDrop ND-1000 is a Shareware software in the category Miscellaneous developed by NanoDrop Technologies.

The latest version of NanoDrop ND-1000 is currently unknown. It was initially added to our database on 06/12/2008.

NanoDrop ND-1000 runs on the following operating systems: Windows.

NanoDrop ND-1000 has not been rated by our users yet.

Write a review for NanoDrop ND-1000!

08/12/2021 SmartGit 21.1.1
08/12/2021 NoteZilla 9.0.6
08/12/2021 System Info for Windows 2021.11.5.0811
08/12/2021 ActiveXperts Network Monitor 870.1082
08/12/2021 BlueStacks App Player 5.2.100.1047
08/02/2021 New Chrome version 92.0.4515.131 available
07/23/2021 Oracle closes more than 300 security vulnerabilities
07/22/2021 Firefox 90.0.1 update available
07/21/2021 Adobe fixes critical vulnerability in Photoshop
07/21/2021 Update to Chrome 92 available
  • » nanodrop 1000 ドライバー
  • » nanodrop software nd1000
  • » nanodrop アップデート nd-1000
  • » nanodrop nd 1000 software download
  • » nd 1000 nanodrop
  • » nano drop nd1000
  • » naodrop nd-1000 download
  • » nanodrop nd 1000
  • » nanodrop nd1000 windows 10 software
  • » nano drop 1000 driver download
NanoDrop ND-1000 spectrophotometer

Websites

  • General FAQ http://nanodrop.com/library/ND%201000%20FAQ.pdf
  • User Manual http://nanodrop.com/Support.aspx?Type=User%20Guides&Cat=NanoDrop%201000
  • Reproducibility Performance Data http://www.nanodrop.com/library/NanoDrop%201000-Reproducibility-Performance-Data.pdf
  • Linearity Performance Data http://www.nanodrop.com/library/NanoDrop%201000-linearity-Performance-Data.pdf
  • Homogeneity Performance Data http://www.nanodrop.com/library/NanoDrop%201000-Homogeneity-Performance-Data.pdf
  • NanoDrop Website http://www.nanodrop.com

Procedure

Typical DNA peak as measured by the Nanodrop on PC
  1. Before starting the software module, clean the sample surfaces with DI water to remove any dried sample that might be present.
    • Alternatively, you can clean the sample surfaces with a Kimwipe moistened with 70% ethanol.
  2. Open the Nanodrop program and the appropriate module (e.g., DNA).
  3. Wipe off the top and bottom sensors of the instrument with a Kimwipe. These are just the polished ends of fiber optic cable, so wiping is sufficient to prevent carryover.
  4. Pipette 2 μL of DD water (stored in a blue microtube) onto the sensor. Bring down the lever arm.
  5. Follow the onscreen prompts to calibrate.
  6. Wipe the sensors and pipette on 2 μL of the corresponding blank (Buffer EB or whatever solution your prep is in). Bring down the lever arm.
  7. Follow the onscreen prompts to blank.
  8. Wipe the sensors and pipette on 2 μL of your sample. Bring down the lever arm.
  9. Click Measure and record the concentration measured.
  10. For DNA, the peak should be at 260 nm, and as a general rule, the 260/280 ratio should be between 1.8 and 2.0.
  11. To test multiple samples, just wipe the sensor in between measurements with a Kimwipe. Recalibration or re-blanking is not necessary.
  12. Clean the sample surfaces once more after you are finished.

Notes

1µl trizol (phenol, chloroform, guanidine) in 1ml H2O

Nanodrop Nd-1000 Software Download

1µl chloroform in 1ml H2O

Measurement volume - 1 vs 2 µl

If your sample volume is very low, you may want to measure 1µl instead of the recommended 2µl. In my hands, however, this doubles the measurement error from about 0.5% to 1% standard deviation/mean. These figures are from repeated measurement of 2 RNA samples, which resulted in 3.5ng/µl SD for a mean of 689ng/µl (2µl volume) vs. 9.2ng/µl SD for a mean of 778ng/µl (1µl), n=5. -- Jakob 2009/7

Liquid column not formed

If you see that a liquid column is not being formed, the surfaces may have been un-conditioned by protein or surfactants. Rub each surface 15-20 times with a dry Kimwipe to buff. Using 2 μL samples should ensure the formation of a column.

Nanodrop Nd 1000 Spectrophotometer

Nanodrop gives strange readings

Occasionally, the Nanodrop can give strange readings. One method to help prevent against this is (after you blank the instrument and wipe the sensor) to pipette on 2 μL of your blank once more and click measure. Then verify that the absorbance is at or near 0 for all wavelengths. If it is not, reclean the sensor and start again. Perhaps this effect is due to a liquid column not being formed?

Nanodrop freezing - cable problems

We had issues of the Nanodrop freezing up sporadically during measurement such that the software became completely unresponsive and the machine was stuck in a measurement state. Tech support responded and said that static discharge can interfere with the computer/machine connection. They had previously sent out a batch of cables which were defective. The USB cable should be white not black. If your USB cable is black, you should probably call them up and ask for a replacement. --Austin 15:17, 18 April 2006 (EDT)

Lab Nanodrop pages

FAQ

Question

How does the Nanodrop convert A260 to concentration in ng/uL?

Answer

From the nanodrop user manual (download at http://nanodrop.com/Support.aspx?Type=User%20Guides&Cat=NanoDrop%201000 )

  • General

The Beer-Lambert equation is used to correlate the calculated absorbance with concentration:
A = E * b * c
Where A is the absorbance represented in absorbance units (A), E is the wavelength-dependent molar absorptivity coefficient (or extinction coefficient) with units of liter/mol-cm, b is the path length in cm, and c is the analyte concentration in moles/liter or molarity (M).

  • For nucleic acid quantification, the Beer-Lambert equation is manipulated to give:

c = A/(E * b)
Where c is the nucleic acid concentration in ng/microliter, A is the absorbance in AU, e is the wavelength-dependent extinction coefficient in ng-cm/microliter and b is the path length in cm. The generally accepted extinction coefficients for nucleic acids are:
• Double-stranded DNA: 50
• Single-stranded DNA: 33
• RNA: 40
For the NanoDrop® ND-1000 Spectrophotometer,paths of 1.0 mm and 0.2 mm are used compared to a standard spectrophotometer using a 10.0 mm path. Thus, the NanoDrop® ND-1000 Spectrophotometer is capable of measuring samples that are 50 times more concentrated than can be measured in a standard spectrophotometer.

Note: absorbance data shown in archive files are represented as displayed on the software screen. For Nucleic Acid, Protein A280 and Proteins and Labels modules, data are normalized to a 1.0 cm (10.0 mm) path. For MicroArray, UV-Vis, Protein BCA, Protein Bradford, Protein Lowry and Cell Culture modules the data are normalized to a 0.1 cm (1.0 mm) path. For high absorbance UV-Vis samples, data are normalized to a 0.1mm path.

Nanodrop

Other References

  • The Biopolymer Calculator (http://paris.chem.yale.edu/extinct.html) calculates the extinction coefficient of nucleic acid sequences using an established base composition and nearest neighbor algorithm.

Comparing concentrations obtained with the calculated extinction coefficient vs. the nanodrop constant:
Calculated concentration was ~20% lower for two RNA sequences (subtillis pheB and BBa_I7101 mRNA) ~~cmc 11:13, 2 Jun 2005 (EDT)

  • http://www.owczarzy.net/extinct.htm For more information on algorithms for calculating extinction coefficients.
  • Warshaw, Ph.D. Thesis, University of California, Berkeley, 1965
  • Warshaw and Tinoco, J. Mol. Biol., 20, 29 (1966)
  • Cantor and Warshaw, Biopolymers, 9, 1059 (1970)
  • Molecular Cloning, Third Edition, A8.19 on Quantitation of Nucleic Acids:

Suggests that for short oligos it is important to calculate the extinction coefficient of a sequence of interest, but that for large molecules the following average extinction coefficients should be used:dsDNA: 50 (uG/ml)-1
ssDNA or RNA: 38 (uG/ml)-1

Question

When you are measuring the concentration of DNA at 260 nm, the softwareautomatically compensates for the fact that you are measuring your sample at a 1 mm path length instead of the standard 1 cm pathlength. For cell culture at OD 600 nm, the software does not do the same thing. Instead it displays the 1mm absorbance. Why?

Answer

This answer was received in an email response from an Application Scientist at Nanodrop Technologies, Inc.

'Yes, the Cell Cultures module is currently displayed at 1 mm and not 10 mm.

Note: absorbance data shown in archive files are represented as displayed onthe software screen. For Nucleic Acid, Protein A280 and Proteins and Labelsmodules, data are normalized to a 1.0 cm (10.0 mm) path. For MicroArray,UV-Vis, Protein BCA, Protein Bradford, Protein Lowry and Cell Culturemodules the data are normalized to a 0.1 cm (1.0 mm) path. For highabsorbance UV-Vis samples, data are normalized to a 0.1mm path.

Regarding microbial cell cultures in suspension:

The short answer: The ND-1000 can be used to estimate bacterial growthcultures, and any limitations will be similar to limitations of otherspectrophotometers.

The long answer: The fundamental issue surrounding bacterial culture growthmeasurements is that an absorbance spectrophotometer is being asked todetermine light scattering caused by particulates in suspension. In thiscase, transmittance is not related to absorbance in the classical sense.Under normal true absorbance conditions, spectrophotometers can becomparable to one another because the sample actually absorbselectromagnetic energy. In the case of a reduction of transmittance causedby light scattering, readings are very dependent on the optics of a specificspectrophotometer as well as the cell type in suspension. For bacterialgrowth determination, even 1cm path length systems can vary greatly due tothe variability of the optics of each system. The ND-1000 will display an'absorbance' value approximately 10 fold less than 1cm systems due to thefact that the instrument is utilizing a 1mm path length. However, thedifference will not be exactly 10 fold due to the reasons I've describedabove. The point is that the differences between the ND-1000 and a'conventional' spec (apart from the 10 fold difference due to path length),will be similar to differences found between spectrophotometers that utilizea 1cm path length.'

Videos

  • Nanodrop it like it's hot! by Brown iGEM